Bruker SkyScan 1174

Technical specification

X-ray Source: Sealed air-cooled x-ray tubes 20 - 50 kV, 40 W
X-Ray Detector: 1280x1024x12bit digital X-ray camera with coupling to P43 scintillator screen by motorized zoom macro lens

Scanning specification

Scanning Area: 6.5 - 30 mm in diameter and length dependent on magnification, 53 - 80 mm total scan length
Scanning and Reconstruction: 1024x1024 or 512x512 pixels in projection and cross section, 16bit depth in acquisition, 32bit float-point reconstruction, 8 or 16bit reconstructed slices
Reconstruction Algorithm: Multithread reconstruction by modified Feldkamp algorithm with volumetric geometry